Answered
create MDD file for Rigol DS 4000 series scope
I suspect you've not installed the hp34401 driver and you are trying to call MAKEMID on that driver. The error message, "The spe...

10 years ago | 0

| accepted

Answered
Can you suggest a way to minimise USB latency using data acquisition tool box with NI6211.
Try using <http://www.mathworks.com/help/daq/ref/inputsinglescan.html inputsinglescan> and <http://www.mathworks.com/help/daq/re...

10 years ago | 0

| accepted

Answered
Instrfindall gives an error in a Matlab executable (and not in Matlab)
Please <http://www.mathworks.com/support/contact_us contact technical support> with the reproduction code.

10 years ago | 0

Answered
Read the outgoing serial data to an IR camera
The serial port functionality in MATLAB is "exclusive" in that only one process can communicate with the serial port - MATLAB or...

10 years ago | 0

| accepted

Answered
What is the difference between using fopen and connect for instrument control? Is there an advantage to using one over the other?
Not all interfaces support FOPEN and CONNECT. You should choose the function to use depending on the kind of object you are crea...

10 years ago | 0

Answered
Can Matlab R2014a, or any other version for 64 bits, on a macbook pro with OS X 10.9.4 use NI-VISA or any other VISA?
Please try the R2015b pre-release or newer versions.

10 years ago | 0

Answered
Instrument Control Toolbox won't recognize NI-VISA drivers on OS X Lion
Have you tried the R2015b pre-release with the latest version of NI VISA for the Mac?

10 years ago | 0

Answered
Input buffer overflow error on HP/Agilent 33120A Arbitrary Waveform Generator
InputBufferSize pertains to the object in MATLAB. It does not control the buffer size of the instrument. I think you want to ch...

10 years ago | 0

| accepted

Answered
How to keep a bluetooth object open in between multiple runs of the same program?
The first two lines of your script: clear all delete(instrfindall); are what is clearing all variables, objects and r...

10 years ago | 0

Answered
Using Simulink to Collect Data from a Network Analyzer
If you already have your MATLAB function that acquires the data from the instrument and formats it for use later, you could cons...

11 years ago | 0

Answered
How to continue GPIB reading with fscanf after EOI?
Have you tried changing the configuration of the <http://www.mathworks.com/help/instrument/eoimode.html EOIMode property> .

11 years ago | 0

Answered
How can I establish communication between Matlab and B&R control system?
Please take a look at the OPC toolbox: http://www.mathworks.com/products/opc/

11 years ago | 0

Answered
bluetooth problem witn instrhwinfo('Bluetooth'); command
Please contact technical support at http://www.mathworks.com/support/contact_us. Remember to provide the full output of executin...

11 years ago | 0

Answered
Instrumental toolbox driver tcpip network analyzer issue
Please contact technical support at http://www.mathworks.com/support/contact_us

11 years ago | 0

Answered
Can Matlab R2014a, or any other version for 64 bits, on a macbook pro with OS X 10.9.4 use NI-VISA or any other VISA?
There are ways to connect to LAN-based instruments without needing to use VISA. Are you mostly connecting to TCPIP instrument...

11 years ago | 0

Answered
IC toolbox, improve speed reading data from spectrum analyzer through the LAN
I strongly suspect that the bottleneck here is not your code, but how fast your instrument is making the measurement and putting...

11 years ago | 0

Answered
Using Thorlabs DPC5500 with Matlab
I think you need to look at the manual for the thorlabs device and see if the commands you are sending to get the instrument to ...

11 years ago | 1

Answered
How to connect an IVI Driver to Instrument Control Toolbox (Test &Measurent Tool) once downloaded on computer?
It looks like the driver is not registered for your configuration in the IVI configuration store. Are you using a 32-bit driver ...

11 years ago | 0

Answered
I²C Restbus Simulation with Instrument Control Toolbox
The I2C functionality in Instrument Control Toolbox is developer to allow your PC to be a I2C master. Contact technical support ...

11 years ago | 0

Answered
How can I communicate between PC and Robot with Bluetooth?
What version of MATLAB are you using? Could you post the output of running 'ver'?

11 years ago | 0

| accepted

Answered
Using Thorlabs DPC5500 with Matlab
Can you point me to the driver you are using? Also if you haven't already, take a look at the documentation on <http://www.math...

11 years ago | 0

Answered
How to transmit data real-time from spectrum analyzer to matlab using the instrument control tool?
Have you looked at <http://www.mathworks.com/help/instrument/examples/reading-inphase-and-quadrature-iq-data-from-a-signal-analy...

11 years ago | 0

Answered
'fclose()' causes matlab to hang, am I missing a toolbox?
Have you tried this with a different instrument (other than your DPO4104)? Also, what version of NI VISA are you using?

11 years ago | 0

Answered
How to save the measurement result(or tracedata) while matlab instrument control tool connecting the Agilent N9020A with IVI drivers
If you're after the IQ data from the spectrum analyzer, take a look at <http://www.mathworks.com/help/instrument/examples/readin...

12 years ago | 0

Answered
How do I send data between two computers using the UDP function from the Instrument Control Toolbox?
The UDP protocol does not guarantee transmission or order of the packets. For most applications, you probably want to use TCPIP...

12 years ago | 0

Answered
MCR 8.2 deployment of UI using instrument control toolbox "Unable to load ni GPIB interface"
Have you checked the <http://www.mathworks.com/support/bugreports/search_results?utf8=%25E2%259C%2593&search_executed=1&keyword=...

12 years ago | 0

Answered
Getting an error "Unable to create VISA object. The required drivers may not be installed properly." in MATLAB using TekVISA version 2.03 to get the data from the RSA 3048B spectrum Analyzer?
Tek VISA 4.0.0 is the current version of Tek VISA. Any reason you are using Tek VISA 2.0? Please upgrade to the latest Tek V...

12 years ago | 0

Answered
How can I get higher throughput using fread with tcpip?
Good to know you've eliminated other potential bottlenecks. We've done performance tests that achieve in excess of 45 *Megaby...

12 years ago | 0

Answered
Instrument Control Toolbox: terminator arriving on input stream but not being detected?
Have you contacted <www.mathworks.com/support technical support> on this issue?

12 years ago | 0

Answered
Can the Total Phase cheetah.dll be used in place of the aardvark.dll for SPI communications?
The DLL with the API to the cheetah adapter does not provide the same entry points, or function signatures. Replacing the DLL wi...

12 years ago | 0